Redefining Cardiovascular Flow Measurement

Overview

|

Features

|

Software

|

Specifications

Where Confidence Meets Measurement Fidelity

Maximum forward instantaneous flow: 100 L/min (1,667 mL/sec)
Maximum reverse instantaneous flow:-100 L/min (-1,667 mL/sec)
Minimum flow:0 L/min
Maximum cardiac output:20 L/min
Flow accuracy:0-50 ml/sec: ±2.5 ml/sec
50-1700 ml/sec: ±4% of measurement
Accuracy of regurgitant volume per ISO-5840:2021:2 ml for 20-200 BPM
Accuracy of regurgitant flow per ISO-5840:10% of 1.5 L/min
Noise Level:<1.5 ml/sec
Flow zeroing interval:Over 30 mins
Beat rate range:20 to 200 BPM pulsatile
Measurement bandwidth:30 Hz to 100 Hz, user-selectable
Typical audible noise level:35 dBa (Normal Working Condition), 45 dBa (During Warm-Up)
Zeroing available on:D-FMS processing module, D-FMS App, In ViVitest
Temperature sensor location:Integrated – embedded in Flow Lumen
Probe construction:Splash-resistant
Fluid temperature range:18 °C to 45 °C
Fluid temperature accuracy:±0.5 °C
Typical fluid temperature reading noise floor:0.1 °C
Maximum fluid pressure:500 mmHg
Probe fluid compatibility:Saline, PBS, Glycerin-Saline blood analog (3-4 cP), other conductive fluids (customization only).
Flow calibration:Automated with SuperPump, Manual Gain Adjustment
Temperature calibration:Manual Offset Adjustment
Probe warm up time:<30 minutes
User-selectable voltage scaling:Up to +/-10 V
Analog output update rate:600 Hz
Analog output latency:Less than 10 ms
Output connector:BNC, cable is provided
Default mode to match ViVitest inputs:±5 V = ± 830 ml/s
Software compatibility• Digital compatibility with D-FMS App v1
• Digital and analog compatible with ViVitest v5
• Analog flow compatible with ViVitest v4, QCTest v3
Laptop requirements:Minimum 1 available USB port compatible,
Windows 10 or 11, RAM 8 GB or greater, I5 processor or greater
Probe dimensions:• Inside Diameter (flow path): 28 mm (no visual obstruction through bore to facilitate imaging)
• Thickness (flow direction): 22 mm
• Total Width: 52 mm
• Total Length: 111 mm
• Weight: 150 g
• Cable Length: 1.5 m
Barb fittings dimensions:25 mm, 30 mm, 35 mm and 1/’2”
Processing module dimensions:Dimensions (H x W x D): 50 mm x 150 mm x 225 mm
System :
System Weight:2 kg
Voltage:100-250 V~
Max power:65 W
Input/Output:Flow probe circular push-pull connector, USB-C connector, BNC analog flow out ±10 V user adjustable, BNC analog aux in/out, Power cord.
Input frequency:50 or 60 Hz
ViVitro Labs Flow Measurement System Part and Components for Mock Flow Loop (MCL) applications
ViVitro Labs Flow Measurement System Part and Components for Mock Flow Loop (MCL) applications
ViVitro Labs Flow Measurement System - 28 mm probe connected to a Mock Circulatory Loop (MCL)
ViVitro Labs Flow Measurement System – 28 mm probe connected to a Mock Circulatory Loop (MCL)
ViVitro Labs Flow Measurement System Processing Module - Input and Output
ViVitro Labs Flow Measurement System Processing Module – Input and Output