Heart Valve Durability Tester

ViVitro
Heart Valve
Durability Tester

Accelerated wear testing
for replacement heart valves
Test module weight
Empty = 20.6 kg/module Full = 22.1 kg/module
Full Controller and Data Acquisition unit weight
8.4kg
Footprint - 6 module configuration
Min Bench Length = 1650 mm* Min Bench Depth = 600 mm Height = 362 mm (without Camera) Height = 400 mm (with Camera)
Footprint - 2 module configuration
Min Bench Length = 850 mm* Min Bench Depth = 600 mm Height = 362 mm (without Camera) Height = 400 mm (with Camera)
Controller power supply requirements
fully loaded 675 W w/ 6 chambers with controller, 1 light ring, heaters on, at 100 mmHg No DAQ, No PC
System operating frequencies
1 Hz to 70 Hz Test sample behavior and frequency response dependant.
Test chamber diameter
Max support OD: 95 mm
Test chamber height & clearance
Total Height/clearance up to 113.5 mm
Piston max stroke volume
38 ml/stroke
Test chamber transparency
Inflow, outflow, left and right, 1 angle at right 45°
Average hydrostatic pressure head
10 mmHg
*Minimal distance between units: 30mm