How many probes can be connected to the flowmeter?

Technically, any I/O channels can be used as input for flow instead of pressure.  This would allow any user to connect more flowmeter controllers.

The ideal wiring configuration of the system setup is shown in the Pulse Duplicator user manual Section 3.4.1 Electrical Connection Diagram.

However, ViVitest is hard coded and the channels in ViVitest are not user defined. Therefore the users will not be able to modify the channel selection.  ViVitest would still believe that the Flow Meter signal is a pressure signal and would try to apply a pressure calibration menu.

Our current electromagnetic flowmeter controller can host 1 single flow probe per unit

111