Corrosion Galvanic

Evaluation of corrosion of device with different type of metals.

Test Procedure

Test specimens should be electrically joined before exposure. External electrical connection should be employed such as to allow current measurement. The coupled assembly is next immersed in the test electrolyte for the period of exposure. Data recorded during exposures may include galvanic current measurements and couple and control specimen potentials measured relative to a suitable reference.

Data are converted into corrosion rate in penetration per unit time (for example, millimetres per year). Rates for both control (uncoupled) and coupled samples should be reported, with the change in rate due to the coupling reported as either the difference between control and coupled samples or as the coupled rate divided by the control rate (acceleration factor).

If required, an optical microscopic metallographic inspection of the sample at the end of the corrosion test and a final SEM inspection could be performed to detect any localized corrosion phenomenon.

Measured Parameters

  • Rest potential before and after the experiments
  • Galvanic current (in A) and galvanic corrosion evolution in function of time
  • Corrosion rate using equation (mm/year)

Scope

Device with different type of metals (for example a base stent material with added marker bands)

Service Levels

Applicable Standards

  • ASTM – F3044 – 20
    Standard Test Method for Evaluating the Potential for Galvanic Corrosion for Medical Implants

Related Tests

OLE1111


ViVitro Labs Advantages

Many companies look to academia to handle their testing needs due to the reduced cost. With ViVitro Laboratory Services you gain the following advantages:

Time to Market
We understand how critical your timelines are. Our expertise and experience ensure your schedule is met and our commitments are delivered on time.

ISO 17025 Accredited
For studies used for regulatory approval this is a must. Accredited by A2LA and Cofrac.

Free Initial Consultations
All of our studies are scoped via no-cost, no obligation, collaborative consultation.

Guaranteed Confidentiality
We happily provide Non-Disclosure Agreements for all services, including initial consultation.

Join over 5000 cardiovascular device professionals who receive our testing and engineering insights and tips every month.