Crossing Profile
A measurement of the maximum diameter over the length of the delivery system, from the proximal end of the mounted stent to the distal tip.
Test Procedure
With the stent pre-mounted or hand crimped on the delivery system in accordance with the instructions for use, and the sheath (if any) in place, measure the maximum diameter over the length from the proximal end of the mounted stent to the distal tip of the delivery system.
Equipment
The measurement instrument is a non-contacting calibrated optical system.
Measured Parameters
Diameter of the length of the device.
Service Levels
Applicable Standards
ASTM F2081 – 06(2017)
Standard Guide for Characterization and Presentation of the Dimensional Attributes of Vascular StentsISO 10555-5:2013
Intravascular catheters — Sterile and single-use catheters — Part 5: Over-needle peripheral cathetersISO 25539-2:2020
Cardiovascular implants — Endovascular devices — Part 2: Vascular stents