Migration Resistance – Fixation Effectiveness

Force required to displace the prosthesis within a mock artery.

Test Procedure

The endovascular prosthesis is deployed into the mock artery. Using a tensile strength machine a uniform rate of separation the endovascular prosthesis is pulled out from the mock artery and the peak force recorded.

Scope

  • Any device using vascular prostheses
  • Vascular prosthesis
  • Biological, synthetic textile/non-textile materials, non-textile material
  • Stent Grafts: SX or BX frame, NiTi, CrCo, SS, Biological tissues, synthetic textile materials/* non-textile material, non-textile material
  • Covered stent: SX or BX frame, NiTi, CrCo, SS, Biological tissue, synthetic textile/non-textile * material, non-textile material
  • Tubular, bifurcated, branched

Service Levels

Applicable Standards

  • ISO 5840-3:2021
    Cardiovascular implants — Cardiac valve prostheses — Part 3: Heart valve substitutes implanted by transcatheter techniques

Related Tests

OLE1101


ViVitro Labs Advantages

Many companies look to academia to handle their testing needs due to the reduced cost. With ViVitro Laboratory Services you gain the following advantages:

Time to Market
We understand how critical your timelines are. Our expertise and experience ensure your schedule is met and our commitments are delivered on time.

ISO 17025 Accredited
For studies used for regulatory approval this is a must. Victoria and Marseille labs accredited by A2LA.

Free Initial Consultations
All of our studies are scoped via no-cost, no obligation, collaborative consultation.

Guaranteed Confidentiality
We happily provide Non-Disclosure Agreements for all services, including initial consultation.

Join over 5000 cardiovascular device professionals who receive our testing and engineering insights and tips every month.