Profile Effect / Flaring
Ensure the distal tip of crimped stent remains in contact with the balloon at all times, while navigating through a tortuous path. The degree of flaring of the stent away from the balloon is measured.
Test Procedure
Uses a contactless optical measurement system to measure the outside diameter of the stent and the outside diameter of the balloon, in a bent configuration. The measured gap between these measurements is the amount of flaring.
Equipment
Contactless optical measurement system
Calibrated gauge blocks
Catheter support fixture
Service Levels
Applicable Standards
ISO 25539-2:2020
Cardiovascular implants — Endovascular devices — Part 2: Vascular stentsFDA-1545
FDA Guidance: Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems – Guidance for Industry and FDA Staff
Related Tests
OLE1209
ViVitro Labs Advantages
Many companies look to academia to handle their testing needs due to the reduced cost. With ViVitro Laboratory Services you gain the following advantages:
Time to Market
We understand how critical your timelines are.
Our expertise and experience ensure your schedule is met and our commitments are delivered on time.
ISO 17025 Accredited
For studies used for regulatory approval this is a must. Victoria and Marseille labs accredited by A2LA.
Free Initial Consultations
All of our studies are scoped via no-cost, no obligation, collaborative consultation.
Guaranteed Confidentiality
We happily provide Non-Disclosure Agreements for all services, including initial consultation.