Strut and Bridge Thickness

Strut and bridge thicknesses of the expanded finished stent are measured by direct measurement using a calibrated optical system. Strut and bridge thickness are measured at the stent mid-length and near each end, at two circumferential locations at each axial location. If strut and bridge thicknesses are systematically different by design, they will be measured and reported separately. Direct measurement may require test samples destruction by cutting, or flattening, or both.

Service Levels

Applicable Standards

  • ISO 25539-2:2020
    Cardiovascular implants — Endovascular devices — Part 2: Vascular stents