Pulse Duplicator
The world’s most
widely-used heart model

- Flow and pressure readings can be collected and analyzed in real-time, meeting ISO 5840 requirements
- Controls pulsatile fluid flows to simulate various cardiac flow conditions including arrhythmia, normal, hypo and hypertensive states at an assortment of cardiac outputs and beat rates
- Transducer sites allow measurement of wall pressures in the atrium, ventricle, and aortic outflow tract
- Flow can be measured directly at the aortic or mitral valve positions
- Flow can be measured directly at the pulmonary and tricuspid valve positions
- Simulates cardiovascular peripheral resistance and compliance
- Thin acrylic windows permit echo imaging, Doppler flow measurements, and PIV for interrogation of flow field regions
- Allows hydrodynamic testing of heart valve replacements in order to assess mechanical and hemodynamic performance
- Ventricle membrane to replicate the natural chamber flow
- Accessories are available for testing other implantable cardiac devices including Occluder and Closure Devices.
Testing Capabilities and Applications
- Mechanical, polymeric and biological surgical valve prosthesis
- Polymeric and biological valve prosthesis for Transcatheter Aortic Valve Replacement (TAVR)
- Polymeric and biological valve prosthesis for Transcatheter Aortic Valve Implantation (TAVI)
- Polymeric and biological valve prosthesis for Transcatheter Tricuspid Valve Replacement (TTVR)
- Polymeric and biological valve prosthesis for Transcatheter Mitral Valve Replacement (TMVR)
- Stentless valves
- Structural heart and peripheral vascular devices
- Occluder and Closure Devices (with customization)
- Evaluate LVAD or other assist devices (with customization)
- Advanced cardiac flow loop systems for pressure and flow monitoring devices
| Pump & MLH | 53 x 92 x 45 cm |
| DAS & Flow System | 48 x 33 x 32 cm |
| Dry weight: Actuator +Model Heart (Left or Right) + VIA | 18 kg |
| Dry weight:Flow Measuring System | 5.6 kg |
| Operating Voltage | 90-240 VAC / 50-60Hz |
| Temperature range | Ambient to 40°C ± 1.0°C |
| Flow rate | 0 – 15 L/min |
| Heart rate | 3 – 200 bpm |
| Valve size Aortic | Up to 035mm |
| Valve size Mitral | Up to 035mm (*070mm with Large Atrium Accessory) |
| Power Consumption | |
| SuperPump pulsatile pump | Max 300W |
| AmPack | Max 10W |
| Heat Bath* | Heater Capacity 1.2kW |
| Laptop | ~60W |
| Flow Meter* | 25W |
*We recommend connecting the Flow Meter to a different circuit than the Heat Bath. The Flow Meter is a sensitive measurement instrument and the large Heat Bath current draw can cause minor disturbances in the flow measurements when the heater switches on.