Balloon Inflation Deflation Time

The time to inflate and deflate a balloon catheter to a predefined target pressure is measured.

Test Procedure

The pressure signal is recorded at a predefined high frequency and is used to estimate the inflation and deflation time at a predefined target pressure.  Between the inflation and deflation phase, the target pressure is held constant for a specific interval of time.

Equipment

The test is performed with an automatic inflation/deflation system.

Scope

  • Angioplasty balloon catheters
  • PTA, PTCA balloon catheters
  • Any interventional systems with balloon technology
  • Any delivery system with balloon technology

Service Levels

Applicable Standards

  • ISO 10555-1:2013
    Intravascular catheters — Sterile and single-use catheters — Part 1: General requirements

  • ISO 25539-2:2020
    Cardiovascular implants — Endovascular devices — Part 2: Vascular stents

  • FDA-1545
    FDA Guidance: Non-Clinical Engineering Tests and Recommended Labeling for Intravascular Stents and Associated Delivery Systems – Guidance for Industry and FDA Staff

Related Tests

OLE1003


ViVitro Labs Advantages

Many companies look to academia to handle their testing needs due to the reduced cost. With ViVitro Laboratory Services you gain the following advantages:

Time to Market
We understand how critical your timelines are. Our expertise and experience ensure your schedule is met and our commitments are delivered on time.

ISO 17025 Accredited
For studies used for regulatory approval this is a must. Victoria and Marseille labs accredited by A2LA.

Free Initial Consultations
All of our studies are scoped via no-cost, no obligation, collaborative consultation.

Guaranteed Confidentiality
We happily provide Non-Disclosure Agreements for all services, including initial consultation.

Join over 5000 cardiovascular device professionals who receive our testing and engineering insights and tips every month.