David Meier, Mariama Akodad, Andrew G. Chatfield, Georg Lutter, Thomas Puehler, Lars Søndergaard, David A. Wood, John G. Webb, Stephanie L. Sellers, and Janarthanan Sathananthan, J Am Coll Cardiol Intv. 2022 Aug, 15 (15) 1532–1539
The aim of this study was to assess the impact of commissural misalignment (CMA) on hydrodynamic function following VIV intervention with the ACURATE neo THV using a bench model. In VIV using the ACURATE neo THV at deep implantation, increasing degree of CMA was associated with THV leaflet interaction with those of the surgical valve and worsening regurgitant fraction. THV leaflet interaction was prevented when there was CA.
Other Products Cited: Heart Valve Testing Pulse Duplicator
Visit SourceK.Yu. Klyshnikov, E.A. Ovcharenko, Yu.A. Kudryavtseva, L.S. Barbarash, RUSSIAN JOURNAL OF TRANSPLANTOLOGY AND ARTIFICIAL ORGANS, Vol. XXII, No 2–2020
The aim of the study was an in vitro hydrodynamic study of the developed prosthetic heart valve of the second generation, designed to carry out an implantation using “valve-in-valve” method. ...Prototypes of the developed prosthesis were studied under simulated physiological conditions of the heart using a Vivitro Labs pulse duplicator (Canada) in a comparative aspect with “UniLine” clinical commercial aortic valve bioprosthesis (Russia)... ...The design of the prosthesis under development has shown its consistency in terms of functional characteristics, both in comparison with the original UniLine prosthesis and with literature data....
Other Products Cited: Heart Valve Testing Pulse Duplicator
Visit SourceEmily N. Boice, David Berard, Jose M. Gonzalez, Sofia I. Hernandez Torres, Zechariah J. Knowlton, Guy Avital, Eric J. Snider, Bioengineering 2022, 9(7), 319
Central vascular access (CVA) may be critical for trauma care and stabilizing the casualty. However, it requires skilled personnel, often unavailable during remote medical situations and combat casualty care scenarios. Automated CVA medical devices have the potential to make life-saving therapeutics available in these resource-limited scenarios, but they must be properly designed. Unfortunately, currently available tissue phantoms are inadequate for this use, resulting in delayed product development. ...A SuperPump (ViVitro Labs, Victoria, BC, Canada) was used for generating pulsatile, cardiac mimicking arterial flow within the flow loop test platform (Figure 1D)... ...In conclusion, the tissue phantom model presented addresses a current testing gap for developing central vascular access medical devices. Troubleshooting is essential and having to rely on live, or even ex vivo, animal testing for evaluating a medical device with physiological mimicking fluid flow or pressure levels will slow device development...
Other Products Cited: Endovascular Simulator Pulsatile Flow Simulation
Visit SourcePaulo Yu, Vibhav Durgesh, Experimental Thermal and Fluid Science, Vol 138, October 1 2022
Aneurysms are abnormal expansion of weakened blood vessels which can be debilitating or fatal upon rupture. Previous studies have shown that hemodynamics play a role in aneurysm formation, growth and rupture. Therefore, the objective of this work is to investigate the impact of flow structures on vortex strength, impinging location, and wall shear stress for two aneurysm models. ...The inflow parameters used in this study, and , were controlled using a ViVitro Labs Inc. SuperPump system... ...The flow evolution results showed that impacted the vortex formation, growth, and dissipation processes for the studied aneurysm model...
Other Products Cited: Analysis and Computational Modeling Flow Visualization Pulsatile Flow Simulation Stent and Stent/Graft SuperPump
Visit SourceLibby H. Welborn, Anna K. Himes, Ida E. Greenlee, Nyna J. DeWitt, Ava T. Burgess, Brandon K. Eberl, Olga Pierrakos, 2022 Systems and Information Engineering Design Symposium (SIEDS)
The aim of this paper is to: 1) showcase the design process of a quadleaflet ePTFE valved conduit with a diameter of 16mm or less and 2) offer a hemodynamic performance comparison. ...The proposed valve was tested using the ViVitro Pulse Duplicator under physiologic cardiac conditions... ...A comparison between the two ePTFE handmade valves (trileaflet and quadleaflet) demonstrated hemodynamic differences in regurgitation due to leaflet number. The handmade valves both showed increased regurgitation compared to a Carbomedics valve (commercially available design). Regurgitation had varying effects on pressure gradients and cardiac output...
Other Products Cited: Heart Valve Testing Pulse Duplicator
Visit SourceK.Yu. Klyshnikov, P.S. Onischenko, Е.А. Ovcharenko, СТМ, 2022, vol. 14, No.2
The aim of the study was to study the complex biomechanics of the aortic valve prosthesis and to analyze the effect of frame mobility on the stress-strain state and geometry of the valve leaflet apparatus using a numerical simulation method, which reproduces the qualitative and quantitative results of its bench tests. ...The simulation was validated by comparing the results of numerical and bench simulation on the ViVitro Labs hydrodynamic system (ViVitro Labs Inc., Canada)... ...The results of the UniLine prosthesis biomechanics modeling have demonstrated that mobility of the frame struts of the heart valve prosthesis is a crucial aspect in assessing the effectiveness of the leaflet apparatus geometry and may be a negative factor in case of a highly elastic leaflet material...
Other Products Cited: Analysis and Computational Modeling Pulse Duplicator
Visit SourcePaulo Yu, Vibhav Durgesh, Fluids, 2022, 7, 165.
Aneurysms are localized expansions of blood vessels which can be fatal upon rupture. Studies have shown that aneurysm flows exhibit complex flow phenomena which consist of single or multiple vortical structures that move within the flow cycle. ...Rep and α were the inflow parameters used to match and simulate the conditions in the human circulatory system. These inflow parameters were precisely controlled with a ViVitro Labs Inc. SuperPump system... ...Both POD and DMD are capable analysis tools to investigate the spatial and temporal behavior of large-scale structures of a given flow field. POD and DMD can be used to create mathematical models to look into the flow evolution in sufficiently high detail and at any time step...
Other Products Cited: Flow Visualization SuperPump
Visit SourceSalwa B. Anam, Brandon J. Kovarovic, Ram P. Ghosh, Matteo Bianchi, Ashraf Hamdan, Rami Haj-Ali & Danny Bluestein - Cardiovascular Engineering and Technology (2022)
Bicuspid aortic valve (BAV) is the most common congenital cardiac malformation, which had been treated off-label by transcatheter aortic valve replacement (TAVR) procedure for several years, until its recent approval by the Food and Drug Administration (FDA) and Conformité Européenne (CE) to treat BAVs. ...Materials and Methods Patient-specific in silico models and 3D printed replicas of 3 BAV patients with different degrees of post-TAVR paravalvular leakage (PVL) were created. Patient-specific TAVR device deployment was modeled in silico and in vitro—following … ...PVL degree and locations were consistent with the clinical data. Cross-validation comparing the stent deformation and the flow parameters between the in silico and the in vitro models demonstrated good agreement....
Other Products Cited: Pulsatile Flow Simulation Pulse Duplicator
Visit SourceL. van de Velde, E. Groot Jebbink, B. A. Zambrano, M. Versluis, J. Tessarek & M. M. P. J. Reijnen: CardioVascular and Interventional Radiology (2022)
...To assess this flow diverting strategy for popliteal aneurysms, flow profiles and thrombus formation likelihood were investigated in popliteal aneurysm models... ...The model was placed in a circulating flow set-up with a blood-mimicking fluid (Fig. 3). At the inlet, a hydraulic piston pump (SuperPump, ViVitro labs, Victoria, Canada) was used to enforce time-varying forward flow...
Other Products Cited: Pulsatile Flow Simulation Stent and Stent/Graft SuperPump
Visit SourcePaulo Yu, Vibhav Durgesh, Fluids, 2022
...Rep and α were two inflow parameters used in this study, and these were controlled using a ViVitro Labs Inc. (Victoria, BC, Canada) SuperPump system...POD provides a method for analyzing complex flow phenomena in great detail. It can be used to study a variety of unsteady, physiological flows to understand the behavior of energetic flow structures..
Other Products Cited: Flow Visualization SuperPump
Visit SourceH. Appa, K. Park, D. Bezuidenhout, B. van Breda, B. de Jongh, J. de Villiers, R. Chacko, J. Scherman, C. Ofoegbu, J. Swanevelder, M. Cousins, P. Human, R. Smith, F. Vogt, B, K. Podesser, C. Schmitz, L. Conradi, H. Treede, H. Schröfel, T. Fischlein, M. Grabenwöger, X. Luo, H. Coombes, S. Matskeplishvili, D. F. Williams, P. Zilla, Front. Cardiovasc. Med., 03 March 2022
...Hydrodynamic testing to determine gradients (ΔP), effective orifice areas (EOA) and regurgitant fractions was performed in a pulse duplicator (Cardiac Output: 5 L/min; 37°C; 70 bpm; Stroke Volume: 32 ± 5 ml; Systolic Phase Duration of 35 ± 5%) (ViVitro Labs Inc; Victoria, BC, Canada)...The BE TAVR system met all requirements for which it was developed...
Other Products Cited: Heart Valve Testing Pulse Duplicator
Visit SourceY. Zhu, A. M. Imbrie-Moore, R. J. Wilkerson, M. J. Paulsen, M. H. Park & Y. Joseph Woo, BMC Cardiovascular Disorders volume 22, Article number: 73 (2022)
...The left heart simulator, which has been previously described, features a pulsatile linear piston pump (ViVitro Superpump, ViVitro Labs, Victoria, British Columbia, Canada) to generate physiologic parameters using the pump controller and software (ViVitest Software, ViVitro Labs) programmed in accordance with ISO 5840 standards for in vitro valve testing...The mitral annular dilation system successfully generated an ex vivo mitral annular dilation model with MR...
Other Products Cited: Ex ViVo Simulation Heart Valve Repair Pulse Duplicator
Visit SourceN Norris, K Lewis, C Settanni, T Becker, A Ducruet, Journal of NeuroInterventional Surgery 2021, 13 (Suppl 1)
...Physiologically-relevant patient flow parameters were tuned within the system using a Super Pump AR (ViVitro Labs)...This in-vitro aneurysm flow model utilized the latest in UV-cured 3D-printing techniques to provide a realistic simulation of neurovascular tortuosity and aneurysm device delivery...
Other Products Cited: Pulsatile Flow Simulation SuperPump
Visit SourceM. Alizadeh, M. Cote, A. Branzan Albu, Proceedings of the IEEE/CVF Winter Conference on Applications of Computer Vision 2022
...Our experimental setup to collect high-quality videos and pressure and velocity data involves tricupsid BHVs mounted in the Pulse Duplicator device from ViVitro Labs...Experiments on four cases that include three different tricuspid BHVs yielded promising results, with the DTW scores showing a good coherence with respect to the simulations...
Other Products Cited: Heart Valve Testing Pulse Duplicator
Visit SourceM. Marin-Cuartas, Y. Zhu, A. M. Imbrie-Moore, M. H. Park, R. J. Wilkerson, M. Leipzig, P. K. Pandya, M. J. Paulsen, M. A. Borger, Y. J. Woo, Interactive CardioVascular and Thoracic Surgery (2021) 1–9
...The severity of acute papillary muscle (PM) rupture varies according to the extent and site of the rupture. However, the haemodynamic effects of different rupture variations are still poorly understood...A 3D printer (Carbon 3D, Redwood City, CA, USA) was used to prototype a modular left heart (left atrium and ventricle) and was mounted to a pulsatile linear piston pump (ViVitro Superpump; ViVitro Labs, Victoria, BC, Canada)...
Other Products Cited: Customized Circulatory Loop Heart Valve Repair SuperPump
Visit SourceA. Imbrie-Moore, Y. Zhu, T. Bandy-Vizcaino, M. Park, R. Wilkerson, Y. Woo, Annals of Biomedical Engineering, 49
Driven by a pulsatile linear piston pump (ViVitro Superpump, ViVitro Labs, Victoria, BC, Canada) which generates physiologic hemodynamics in accordance with ISO 5840 in vitro cardiac valve testing standards.
Other Products Cited: Heart Valve Repair SuperPump
Visit SourceB. Ncho, A. Siefert, V. Sadri, J. Ortner, A. Yoganathan. Journal or Med. Devices
The physiologic pressures and flows were achieved by adjustment of systemic resistance and compliance elements, and flow driven from a piston pump (Vivitro Labs Inc. Victoria, Canada).
Other Products Cited: Flow Visualization Heart Valve Testing SuperPump
Visit SourceP. Yu, V. Durgesh, Journal of Biomechanical Engineering, 5
The inflow conditions are precisely controlled using a ViVitro SuperPump system where nondimensional fluid parameters such as peak Reynolds number (Rep) and Womersely number (α) are varied from 50–270 and 2–5, respectively.
Other Products Cited: Flow Visualization SuperPump
Visit SourceJ. Westenberg, H. van Assen, P. van den Boogaard, J. Goeman, H. Saaid, J. Voorneveld, J. Bosch, S. Kenjeres, T. Claessens, P. Garg, M. Kouwenhoven, H. Lamb, Magnetic Resonance in Medicine
A pulsatile hydraulic piston pump (Vivitro Labs, Victoria, Canada) was connected to the tank with semirigid tubing. This pump, positioned outside the scanner room, generated a sinusoidal-like pulsatile waveform with 70-bpm frequency without net flow, resulting in a pressure waveform inside the tank that created contraction of the filled LV phantom with a 300-ms systolic period and 50-mL stroke volume.
Other Products Cited: Flow Visualization SuperPump
Visit SourceR. Whiting, E. Sander, C. Conway, T. J. Vaughan. 2021.
...To evaluate the hydrodynamic performance of each valve, in vitro bench testing was performed using a Vivitro Pulse Duplicator (Figure 2a) (Vivitro Labs, Inc. Victoria, B.C.) according to ISO 5840...
Other Products Cited: Pulsatile Flow Simulation Pulse Duplicator